Chapter 6: Routine Operation<br />
Note: If the rotor cover is not pulled over the rotor, the instrument will not function.<br />
Note: It is recommended you change the dispensing cassette or tubing set after<br />
dispensing about 3000 – 5000 96-well plates, depending on the dispensing frequency and<br />
the liquids used.<br />
Note: Keep the dispensing cassette in the rest position when the dispensing cassette is<br />
not in use.<br />
6.4 How to insert and adjust the plate adapter<br />
The Multidrop <strong>384</strong> requires a plate adapter for the plate carrier to function. Two different plate<br />
adapters are included (Fig. 6.7).<br />
Fig. 6.7 Two different plate adapters<br />
The PLATE ADAPTER, PLATE<br />
HEIGHT MAX. 15 MM, is<br />
suitable for plates approximately<br />
15 mm high, for example, 96well<br />
plates and some <strong>384</strong>-well<br />
plates.<br />
The PLATE ADAPTER, PLATE<br />
HEIGHT MAX. 11 MM, is<br />
suitable for lower plates, for<br />
example, <strong>384</strong>-well plates.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 25