26.11.2012 Views

Thermo Scientific Multidrop® 384 - Lab Equipment, Industrial ...

Thermo Scientific Multidrop® 384 - Lab Equipment, Industrial ...

Thermo Scientific Multidrop® 384 - Lab Equipment, Industrial ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

6.7 How to prime<br />

Chapter 6: Routine Operation<br />

Note: Ensure that the instrument is sufficiently level on the table to prevent liquids from<br />

overspilling when priming.<br />

Note: Ensure that the liquid temperature is close to the calibration temperature of 22°C.<br />

Differences greater than ± 2°C will cause an error in the results.<br />

Note: Silicon tubing (silicon rubber) is resistant to a number of chemical liquids. Most<br />

water solutions typically used in a medical laboratory can be dispensed with the Multidrop<br />

<strong>384</strong>. Some strong substances may, however, harm the silicone tubing and cause swelling<br />

or breaking. If the substance has even a slight effect on silicone, the calibration of the<br />

cassette may be affected and the cassette may become unusable.<br />

1. Place the eight tube ends into the same liquid reservoir or into different reservoirs taking care<br />

that all tube ends are below the liquid level. Use the tubing weight provided if possible.<br />

2. Check that the priming vessel is firmly inserted next to the plate carrier. The Multidrop <strong>384</strong> will<br />

not operate if the priming vessel is not completely inserted into its slot.<br />

3. Press the prime/drop key until all eight channels are dispensing continuously into the priming<br />

vessel and there are no more air bubbles in the tubes. It is recommended you fill the priming<br />

vessel two or three times with the reagent to be used before starting the actual dispensing in<br />

order to bring the tubes into working condition.<br />

6.8 Dispensing<br />

Note: If you press the prime/drop key at the same time as the rotor cover is pulled over<br />

the rotor, the program enters into the volumetric calibration mode.<br />

Note: You can empty the contents of the priming vessel back into the reagent reservoir by<br />

pulling the priming vessel out of its slot if necessary. It is recommended you do this if you<br />

use expensive reagents. Therefore, wash the priming vessel always after each use to<br />

avoid any cross contamination. If you use several liquids at the same time in different rows,<br />

you can order a special eight-cavity priming vessel for this purpose to prevent mixing the<br />

liquids. Also a self-draining (flow-through) priming vessel is available. Refer to section 11.4<br />

Priming vessels (consumables).<br />

6.8.1 How to dispense automatically<br />

1. Set the required reagent volume per well. The permitted volume range is 5 ― 100 µl for <strong>384</strong>-well<br />

plates and 5 ― 395 µl for 96-well plates.<br />

2. Set the required number of columns to be dispensed per plate.<br />

3. Place the microplate onto the plate carrier equipped with a suitable plate adapter. Refer to<br />

section 6.4 How to insert and adjust the plate adapter.<br />

4. Press the prime/drop key to prime the tubes. Refer to section 6.7 How to prime.<br />

<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 32

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!