7.2 How to wash the dispensing cassette<br />

Chapter 7: Maintenance<br />

Wash the tubes by priming them first with tap water and then with deionized distilled water. Make sure<br />

that all the tubes are washed properly.<br />

If cleaning with water is insufficient, use a mild laboratory detergent and then prime with large<br />

amounts of tap water followed by deionized distilled water.<br />

The dispensing cassette can be dried at room temperature.<br />

After washing, store the dispensing cassette in the rest position (Fig. 6.18).<br />

If the instrument is not used for a couple of hours, insert the dispensing cassette into the rest position.<br />

Refer to section 6.8.4 After use. Handle the dispensing cassette with great care so that you do not<br />

damage the tubes and dispensing tips. These factors will affect the service life of the dispensing<br />

cassette.<br />

7.3 How to autoclave the dispensing cassette<br />

The standard tube dispensing cassette (Cat. no. 24072670) can be autoclaved 50 times with the same<br />

tubes.<br />

The autoclaving conditions of the dispensing cassette are as follows: 1 bar pressure at 121°C for<br />

20 minutes.<br />

Note: After autoclaving, the dispensing cassette must cool down to room temperature<br />

before use.<br />