Chapter 7: Maintenance<br />

18. When the tube fasteners have all been adjusted roughly to the middle of the sight strip, close the<br />

calibration screw cover (Fig. 7.34).<br />

Fig. 7.34 Dispensing cassette precalibrated<br />

19. Insert the eight tubes into the tubing weight either at this stage or earlier (Fig. 7.35). The new<br />

tubing set has now been successfully fitted.<br />

Fig. 7.35 Inserting the tubes into the tubing weight<br />

20. Calibrate the dispensing cassette according to the instructions in section 7.5 How to recalibrate<br />

the dispensing cassette.<br />

<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 56