7.6 How to change the tubing set<br />

7.6.1 Changing the tip band<br />

Chapter 7: Maintenance<br />

1. Place the dispensing cassette onto the laboratory bench with the dispensing tips pointing down.<br />

2. Unfasten the four cover retaining screws of the lower part of the dispensing cassette (Fig. 7.11).<br />

Fig. 7.11 Removing the cover plate of the lower part of the dispensing cassette<br />

3. Remove the tip manifold, the tube fasteners and the rest of the tubing from the lower part of the<br />

dispensing cassette. Then detach the tip band from the tubing by pulling the tubes away from the<br />

tip band (Fig. 7.12).<br />

Fig. 7.12 Removing the old tip band<br />

<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 45