Fig. 7.15 New tubing set with the tip band attached<br />

Chapter 7: Maintenance<br />

2. Remove the four cover retaining screws on the lower part of the dispensing cassette, the tubes<br />

with the tube fasteners and the tip manifold with the tips, the three cover retaining screws on the<br />

upper part of the dispensing cassette, the two tension limiting wires, the calibration screws and<br />

the tubes with the tube fasteners. Save the calibration screws and the tubing weight for further<br />

use when installing the new tubing set.<br />

3. Insert the tip manifold so that the tips go through the holes in the lower part of the dispensing<br />

cassette (Fig. 7.16) and guide the tubes through the eight openings in the side wall by pressing<br />

the small cubic tube fasteners into place (Fig. 7.17). Ensure that each tube is in its designated<br />

opening (Fig. 7.18).<br />

Tip band<br />

Tubing identification plate<br />

with tube lot number<br />

Tube fasteners<br />

Fig. 7.16 Inserting the tip manifold into the lower part of the dispensing cassette<br />

<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 47