7.5 How to recalibrate the dispensing cassette<br />

Chapter 7: Maintenance<br />

The dispensing cassette has been calibrated by the manufacturer with deionized distilled water at<br />

22°C ± 2°C. It is recommended you recalibrate the cassette after dispensing about 1000 96-well<br />

plates or after every three months. It is also recommended you recalibrate the dispensing cassette if<br />

you have forgotten the cassette in the Multidrop <strong>384</strong> under tension for prolonged periods, for example,<br />

a weekend or longer.<br />

Two different calibration methods are described below, that is, volumetric calibration and gravimetric<br />

calibration.<br />

7.5.1 Volumetric calibration<br />

For volumetric calibration, use the following steps:<br />

1. Remove the priming vessel and microplate.<br />

2. Install the dispensing cassette into the Multidrop <strong>384</strong> if it is not already installed. Place the<br />

dispensing cassette into the rest position (Fig. 6.18). Refer to section 6.3 How to insert the priming<br />

vessel and the dispensing cassette.<br />

3. Install the tube of the calibration device into the tip of the channel to be recalibrated (Fig. 7.5), for<br />

example, A.<br />

Fig. 7.5 Placing the tube onto the tip of channel A<br />

4. Insert the dispensing cassette into the working position so that the lower and upper parts of the<br />

dispensing cassette are inserted into their slots (Fig. 6.5).<br />

5. Pull the rotor cover over the rotor.<br />

6. Place the tube of channel A into the reservoir with water at room temperature (working<br />

temperature).<br />