Chapter 7: Maintenance<br />

4. Replace the tip band (Fig. 7.13). The tips of the tip band must not be clogged in any way but be<br />

clean to function correctly (Fig. 7.14).<br />

Fig. 7.13 Tip band removed<br />

Insert tube here<br />

Fig. 7.14 New tip band<br />

7.6.2 Fitting the new tubing set step by step<br />

1. First place the new tubing set onto the laboratory bench with the tip band attached (Fig. 7.15).<br />

Ensure that you have saved the eight calibration screws and the tubing weight from the previous<br />

tubing set.<br />

<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 46<br />