Calibration screw<br />
cover opened<br />
Chapter 7: Maintenance<br />
Fig. 7.7 Screwing the calibration screws according to the calibration results<br />
14. Press the empty key and adjust the water level in the pipette to the lower mark and repeat Steps<br />
10 to 13 until the result is 1 ml. Three accurate calibrations must be obtained.<br />
15. Change the tube to the next channel and repeat Steps 10 to 14.<br />
16. After all the channels have been recalibrated, press the stop key to reset the Multidrop <strong>384</strong> into<br />
normal operation and reinstall the priming vessel.<br />
17. Close the calibration screw cover.<br />
7.5.2 Gravimetric calibration<br />
For gravimetric calibration, use the following steps:<br />
1. Install the dispensing cassette into the working position so that the lower and upper parts of the<br />
dispensing cassette are inserted into their slots (Fig. 6.5).<br />
2. Pull the rotor cover over the rotor.<br />
3. Ensure that the priming vessel is installed. Refer to section 6.3 How to insert the priming vessel<br />
and the dispensing cassette.<br />
4. Ensure that you have the plate adapter for 15 mm high plates installed. Refer to section 6.4 How<br />
to insert and adjust the plate adapter.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 42