6.8.4 After use<br />
Chapter 6: Routine Operation<br />
After all the plates have been dispensed, press the empty key to return the reagent or liquid from the<br />
tubes to the reservoir. This saves expensive reagents.<br />
If there is only a short break in dispensing, no washing or cleaning of the tubes is necessary.<br />
However, after pressing the empty key, release the tubing tension by pulling the upper part of the<br />
cassette from its slots and place it into its rest position slots located on the left-hand side of the rotor<br />
(Fig. 6.18).<br />
If there is a longer break (for example, a day), wash the dispensing cassette by priming it with<br />
deionized distilled water or with special washing detergent if necessary. In the end, rinse the washing<br />
detergent away with deionized distilled water.<br />
After washing, store the dispensing cassette in the rest position (Fig. 6.18).<br />
Note: Keep the dispensing cassette in the rest position when the dispensing cassette is<br />
not in use.<br />
Fig. 6.18 Rest position of the dispensing cassette<br />
<strong>Thermo</strong> <strong>Scientific</strong> Multidrop ® <strong>384</strong> User Manual; Rev. 3.4, Cat. no. 1507010 34