<strong>Instruction</strong> Manual pH 1100/ 21003.2.3 Taking pH MeasurementsNOTE: Be sure to remove the electrode soaker bottle orprotective rubber cap on the electrode before measurement.1. Rinse the probe with deionized or distilled water before useto remove any impurities adhering to the probe body. If thepH electrode dehydrated, soak it for 30 minutes in electrodestorage solution or a 2M to 4M KCl solution.2. Press ON. The upper display shows current pH reading andtemperature, and indicates automatic or manualtemperature compensation. The lower display shows month,date, year and time. See Figure 35.7.179pH 25.3°C ATCJan 18 2002 08:46:25Figure 357.546pH 25.4°C ATCJan 18 2002 08:46:30Figure 363. Dip the probe into the sample. Make sure the sensor or theglass bulb of the electrode is completely immersed into thesample. Stir the probe gently to create a homogenoussample.4. Allow time for the reading to stabilize. The pH indicatorblinks when reading is not stable. See Figure 36. When thereadings stabilizes, the pH indicator freezes and no longerblinks.17