<strong>Instruction</strong> Manual pH 1100/ 210010.4.8 Ion calibration data reset setup program P3.6Use this program to erase all ion calibration data.1. Enter ion setup mode as described in steps 1-4 in Section10.4.1 Entering ion setup mode (page 43).2. Use Δ/MI key to scroll to program Ion Cal Reset: P3.6. SeeFigure 155.3. Press ENTER to enter program Ion Cal Reset: P3.6.4. Press >/CAL key to toggle between Ion Reset: Yes andNo. The selection will flash. See Figure 156.5. Press ENTER to confirm selection.• If you selected Yes, the screen flashes Ion beingReset, Please wait... See Figure 157. The meter thenreturns to setup mode.• If you select No, the screen flashes Reset Aborted andthen returns to the setup mode. See Figure 158.This is the last ion setup program. Press ∇/MR to go to theprevious program. Press Δ/MI to go back to the first ionsetup program. Press EXIT to return to mV measurementmode.ModeI on Cal ResetPress ENT. toI on Reset: Yes NoSel ec t & Pr es s ENTERI on being ResetPl e as e wai t . . .Reset Aborted:P3.6SelectFigure 155Figure 156Figure 157Figure 15849