<strong>Instruction</strong> Manual pH 1100/ 2100pH Cal Reset :P1.7Press ENT. t o SelectpH Reset: Yes NoSel ec t & Pr es s ENTERpH being ResetPl e a s e wa i t . . .Reset AbortedFigure 105Figure 106Figure 107Figure 10810.2.9 pH calibration data reset setup program P1.7Use this program to erase all pH calibration data.1. Enter pH/Temperature setup mode as described in steps 1-4 inSection 10.2.1 Entering pH/temperature setup mode (page 30).2. Press Δ/MI to scroll to program pH Cal Reset: P1.7. SeeFigure 105.3. Press ENTER to enter program pH Cal Reset: P1.7.4. Press >/CAL to toggle between pH Reset: Yes and No. Theselection will flash. See Figure 106.5. Press ENTER to confirm selection.• If you selected Yes, the screen flashes pH being Reset,Please wait... See Figure 107. The meter then returns tosetup mode.• If you select No, the screen flashes Reset Aborted andthen returns to the setup mode. See Figure 108.This is the last pH setup program. Press ∇/MR to go to the previousprogram. Press Δ/MI to go back to the first pH setup program. PressEXIT to return to the pH measurement mode.36