<strong>Instruction</strong> Manual pH 1100/ 210010.3.6 mV calibration data reset setup program P2.4Use this program to erase all mV calibration data.1. Enter mV setup mode as described in steps 1-4 in Section10.3.1 Entering mV setup mode (page 37).2. Use Δ/MI key to scroll to program mV Cal Reset: P2.4.See Figure 125.3. Press ENTER to enter program mV Cal Reset: P2.4.4. Press >/CAL key to toggle between mV Reset: Yes and No.The selection will flash. See Figure 126.5. Press ENTER to confirm selection.• If you selected Yes, the screen flashes mV being Reset,Please wait... See Figure 127. The meter then returns tosetup mode.• If you select No, the screen flashes Reset Aborted andthen returns to the setup mode. See Figure 128. This isthe last mV setup program. Press ∇/MR to go to theprevious program. Press Δ/MI to go back to the first mVsetup program. Press EXIT to return to mV measurementmode.mV Cal Reset :P2.4Press ENT. to SelectFigure 125mV Reset: Yes NoSel ec t & Pr es s ENTERmV being ResetPlease wait . . .Reset AbortedFigure 126Figure 127Figure 12842