<strong>Instruction</strong> Manual pH 1100/ 210010.2.6 pH Calibration Due Alarm Setup program P1.4Use this program to set a reminder when the next calibration is due.Choose from 1 to 365 days from last calibration.1. Enter pH/Temperature setup mode as described in steps 1-4 in10.2.1 Entering pH/temperature setup mode (page 30).2. Use Δ/MI to scroll to program pHCalDue Alarm: P1.4. SeeFigure 91.3. Press ENTER to enter program pHCalDue Alarm: P1.4.4. Press >/CAL to toggle between Cal Alarm Due:On and Off. Theselection will flash. See Figure 92.5. Press ENTER to confirm selection.pHCalDue Alar m : P1.4Press ENT. t o SelectFigure 91Ca l Al a r m Du e : On Of fSel ec t & Pr es s ENTERFigure 92Keyin Cal Due Period(1-365 days) = 1Figure 93• If you selected On, continue to step 6• If you select Off, pressing ENTER returns to the setup mode.Press Δ/MI to continue to the next setup program, or press ∇/MRto go to the previous program. Press EXIT to return to the pHmeasurement mode.6. If you selected On in step 5, the display reads Key in Cal DuePeriod (1-365 days) =. Key number of days using the numerickeypad. If you make a mistake, use the Δ/MI or ∇/MR keys tohighlight mistake, then re-enter. See Figure 93.7. Press ENTER to confirm number of days to next calibration,screen flashes Sys Updating, Please wait... and returns to thesetup mode. Press Δ/MI key to continue to the next setupprogram, or press ∇/MR to go to the previous program. PressEXIT to return to the pH measurement mode.33

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!