<strong>Instruction</strong> Manual pH 1100/ 210010.3.3 mV calibration due alarm setup program P2.1Use this program to set a reminder when the next calibration isdue. Choose from 1 to 365 days from last calibration.1. Enter mV setup mode as described in steps 1-4 in Section10.3.1 Entering mV setup mode (page 37).2. Use Δ/MI key to scroll to program mVCalDue Alarm: P2.1.See Figure 116.3. Press ENTER to enter program mVCalDue Alarm: P2.1.4. Press >/CAL to toggle between Cal Alarm Due:On and Off.The selection will flash. See Figure 117.5. Press ENTER to confirm selection.• If you selected On, continue to step 6.• If you select Off, pressing ENTER returns to the setupmode.mVCalDue Alarm :P2.1Pr es s ENT. t o Sel ec tFigure 116Ca l Al a r m Due On Of fSel ec t & Pr es s ENTERFigure 117Keyin Cal Due Period(1-365 days) = 1Figure 118Press Δ/MI to continue to the next setup program, or press∇/MR to go to the previous program. Press EXIT to return tomV measurement mode.6. If you selected On in step 5, the display reads Key in CalDue Period (1-365 days) =. Key number of days using thenumeric keypad. If you make a mistake, use the Δ/MI or∇/MR keys to highlight mistake, then re-enter.See Figure 118.7. Press ENTER to confirm number of days to next calibration,screen flashes Sys Updating, Please wait... and returns tothe setup mode.Press Δ/MI to continue to the next setup program, or press∇/MR to go to the previous program. Press EXIT to return tomV measurement mode.39

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!