<strong>Instruction</strong> Manual pH 1100/ 210010.5.6 Communication setup mode P4.4Use this program to set parameters of serial communication.1. Enter meter setup mode as described in steps 1-4 in Section10.5.1 Entering meter setup mode (page 50).2. Use Δ/MI key to scroll to program Tx/Rx Parameter: P4.4.See Figure 175.3. Press ENTER to enter program Tx/Rx Parameter: P4.4.4. Press >/CALkey to toggle between Baud Rate 4800, 9600 and19200. The selection will flash. See Figure 176. Press ENTERto confirm selection and go to the next setup screen.5. Press >/CALkey to toggle between Stop Bit One and Two.The selection will flash. See Figure 177. Press ENTER toconfirm selection and go to the next setup screen.6. Press >/CALkey to toggle between Parity Bit None, Odd andEven. The selection will flash. See Figure 178.7. Press ENTER to confirm selection and return to the setup mode.Press Δ/MI key to continue to the next setup program, or press∇/MR to go to the previous program. Press EXIT to return tomeasurement mode.Tx/ Rx Parameter:P4.4Press ENT. t o SelectFigure 175Baud Rat e4800 9600 19200Figure 176St op Bit : One TwoFigure 177Par i t y Bi t :None Odd EvenFigure 17810.5.7 Data transfer setup mode P4.51. This program lets you select what data is sent to your printer orcomputer.2. Enter meter setup mode as described in steps 1-4 in Section10.5.1Entering meter setup mode (page 50).3. Use Δ/MI key to scroll to program Data Transfer: P4.5. SeeFigure 179.4. Press ENTER to enter program Data Transfer: P4.5.5. Press >/CALkey to toggle between Data send to printer Currentand Memory. .The selection will flash.See Figure 180.Dat a Tr ansf er :P4.5Press ENT. t o SelectFigure 179Data send to printerCurrent Memor yFigure 1806. Press ENTER to confirm selection and return to the setup mode.Press Δ/MI key to continue to the next setup program, or press∇/MR to go to the previous program. Press EXIT to return tomeasurement mode.53

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!