<strong>Instruction</strong> Manual pH 1100/ 210010.5.8 Memory clear setup mode P4.6Clears all pH, mV and Ion (pH2100 only) memory points storedin memory.1. Enter meter setup mode as described in steps 1-4 in Section10.5.1 Entering meter setup mode (page 50).2. Use Δ/MI key to scroll to program Mem Clear Setup: P4.6.See Figure 181.3. Press ENTER to enter program Mem Clear Setup: P4.6.4. Press >/CAL key to toggle between Memory Clear Yes and No.The selection will flash. See Figure 182.5. Press ENTER to confirm selection.• If you select Yes, the screen flashes Memory beingCleared, Please wait... See Figure 183. The meter thenreturns to the setup mode.Me m Cl e a r Se t u p : P4 .6Press ENT. t o SelectFigure 181Me mo r y Cl e a r : Ye s NoSelect & Press ENTERFigure 182Memory being ClearedPl eas e wai t . . .Figure 183• If you select No, the meter returns to the setup mode.Press Δ/MI to continue to the next setup program, orpress ∇/MR to go to the previous program. Press EXIT toreturn to measurement mode.10.5.9 Meter reset setup mode P4.7Resets the meter to factory default settings:• pH/mV/Ion* calibration data is erased• Manual temperature compensation is set to 25.0°C• pH/mV/Ion* measurement alarm is off• pH/mV/Ion* calibration due alarm is off• Stability indicator is turned on with low stability setting• Data logging and memory recall has manual setting• Backlight is on with 2.0 minute duration• Memory is cleared*Ion features on the CyberScan pH2100 meter only1. Enter meter setup mode as described in steps 1-4 in Section10.5.1 Entering meter setup mode (page 50).2. Use Δ/MI key to scroll to program Meter Reset: P4.7.See Figure 184.3. Press ENTER to enter program Meter Reset: P4.7.4. Press >/CAL key to toggle between Meter Reset Yes and No.The selection will flash. See Figure 185.5. Press ENTER to confirm selection.Meter Reset :P4.7Press ENT. t o SelectMeter Reset: Yes NoSelect & Press ENTERMet er being ResetPl eas e wai t . . .Reset AbortedFigure 184Figure 185Figure 186Figure 187• If you selected Yes, the screen flashes Meter beingReset, Please wait...See Figure 186. The meter thenreturns to measurement mode.• If you select No, the screen flashes Reset Aborted. SeeFigure 187. The meter then returns to the measurementmode.54

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!