Ultrasonic milk analyzerCal 1 Samp LowFat = f.ffSNF=y.yyDen=d.dd------------------------------Lac=l.llSol=s.ssPro=p.ppExitWhich reminds to enter the values <strong>for</strong> the sample with low FAT. Enter thevalues analogically to the procedure, described in the previous sample. Afterentering the last parameter position the cursor on “Exit”, press “Enter” andthe following menu will appear:Recal 1Put sample Low 5timesMake 5 times measurement of the low FAT sample.After 5th measurement <strong>com</strong>pletion automatically appears the menu:Recal 1Put sample waterWhich reminds <strong>for</strong> 5-times water measurement. There is no need to entervalues after placing the sample-holder. Measurement starts directly.After the 5 th measurement appears the menu:RecalibratedAnalyzer readyThis means that the calibration was <strong>com</strong>pleted successfully and the analyzeris recalibrated <strong>for</strong> cow milk, marked as “Calibr 1”.Switch off the power supply of the device and switch it on again.The device is ready to work with the new calibration.If during work the sample’s temperature exceeds the temperature range 15-25° С the following message appears:<strong>Operation</strong> manual 33/98