Ultrasonic milk analyzer7. STARTING THE DEVICE IN A SERVICE TEST/SETUPOPERATIONAL MODE. MENUS DESIGNATION7.1. Starting the device in a service Test/Setup operational mode.In order to start the Setup of the device the operator has to press the buttonEnter and without releasing it to switch on the power supply of the device, towait <strong>for</strong> the starting identification messages and to release the button afterthe following message appears on the display:Release buttonto start setupAfter releasing the button on the display is shown:MA SetupFollowed by possible to be entered by the operator menus:Special modesCorrectionsSettings------------------------------TestspH & Co MeterAccessoriesExitYou may move in the menus by using buttons “up” ▲ and “down” ▼.If by pressing the button Enter you choose a menu, each menu offers newpoints/submenus. When Exit is chosen the device leaves the Setup modeand returns to normal work.<strong>Operation</strong> manual 35/98