No. 42 (PH 8)<br />

Great impact of the brand-new Sysmex XN-Series automated hematology analyzer<br />

on workflow efficiency and utility<br />

Etsuko Hamada, Masato Maekawa<br />

Department of Laboratory Medicine, Hamamatsu University School of Medicine, Hamamatsu, Japan<br />

Introduction:<br />

Recently, higher-quality testing environment has been increasingly required. To counteract these<br />

demands, Sysmex has introduced the XN-Series automated hematology analyzer. Here we report our<br />

evaluation results of the XN-Series basic per<strong>for</strong>mance and the integrated hematology testing system.<br />

Method: (1) Basic per<strong>for</strong>mance: We evaluated precision, linearity, correlation with a previous analyzer<br />

(XE-2100, Sysmex) and correlation with the manual method. (2) Usability: We evaluated the usability<br />

of newly introduced functions of CNA-NET, which include automated smear preparation, selection of<br />

specific sample and sample requiring manual count. Results: (1) Basic per<strong>for</strong>mance: Precision (CBC)<br />

was 0.42-5.21 (CV%), and linearity was confirmed up to WBC 378×10 9 /L, RBC 7.81×10 12 /L, HGB 235<br />

g/L, HCT 72.1%, and PLT 1,691×10 9 /L. The correlation coefficients (r) with the previous analyzer<br />

were CBC 0.998–0.981, WBC classification ratio 0.995-0.917 and reticulocyte ratio 0.978. The<br />

correlation coefficient of WBC classification ratio with the manual method was 0.612-0.987. In the<br />

evaluation using samples with WBC ≤ 0.5×10 9 /L, WBC precision was 7.75-8.21 (CV%) <strong>for</strong> Whole<br />

blood mode and 1.29-4.17 <strong>for</strong> Low WBC mode, and the good correlation with the manual method was<br />

observed. In the evaluation using samples with PLT ≤ 50×10 9 /L, the correlation coefficient between<br />

PLT-I (impedance) and PLT-F (fluorescence) was 0.973, and with the manual method was 0.975 <strong>for</strong><br />

PLT-I and 0.980 <strong>for</strong> PLT-F. (2) Usability: The number of analyzer operator to measure 700 samples/<br />

day decreased from 1.5 to 1 person. Conclusion: Basic per<strong>for</strong>mance of the XN-Series was satisfactory.<br />

Especially, precision and accuracy in the low number of WBC and PLT were excellent. There<strong>for</strong>e it is<br />

possible to report higher clinical value result with the automatic measurement method. Use of the<br />

automated re-testing function and the CNA-NET made reporting time and processing time reduced. The<br />

newly introduced integrated hematology testing system improved the quality of the clinical testing.<br />

- 95 -