9. Alternate trigger (Trigger Mode: Pulse)<br />

Alternate trigger (Trigger Type: Pulse) Menu is shown as Fig. 36.<br />

Fig. 36 Alternate trigger (Trigger Type: Pulse) Menu<br />

Alternate trigger (Trigger Type: Pulse) menu list:<br />

MENU SETTING INSTRUCTION<br />

Alternate<br />

Mode<br />

Pulse<br />

Set vertical channel trigger type as pulse trigger.<br />

Source<br />

CH1<br />

Select CH1 as the trigger source.<br />

CH2<br />

Select CH2 as the trigger source.<br />

AC<br />

DC<br />

Not allow DC portion to pass.<br />

Allow all portion pass.<br />

Coupling<br />

HF<br />

Not allow high frequency of signal pass and only low<br />

frequency portion pass.<br />

LF<br />

Not allow low frequency of signal pass and only high<br />

frequency portion pass.<br />

Polarity<br />

Choose the polarity<br />

when<br />

Select pulse width condition and turn the M knob to set<br />

time.<br />

Mode<br />

Holdoff<br />

Auto<br />

Holdoff<br />

Reset<br />

Acquire waveform even no trigger occurred<br />

100ns~10s, adjust M knob to set time interval before<br />

another trigger occur.<br />

Set Holdoff time as 100ns<br />

Term interpretation<br />

1. Source:<br />

Trigger can occur from several sources: Input channels (CH1, CH2), AC Line, Ext, Ext/5.<br />

* Input: It is the most commonly used trigger source. The channel will work when selected as a<br />

trigger source whatever displayed or not.<br />

* Ext Trig: The instrument can trigger from a third source while acquiring data from CH1 and CH2. For<br />

example, you might want to trigger from an external clock or with a signal from another part<br />

of the test circuit. The Ext, Ext/ 5 trigger sources use the external trigger signal connected to<br />

the EXT TRIG connector. Ext uses the signal directly; it has a trigger level range of +1.6 V to<br />

-1.6 V. The EXT/ 5 trigger source attenuates the signal by 5X, which extends the trigger level<br />

range to +8 V to -8 V. This allows the oscilloscope to trigger on a larger signal.<br />

-122-