Function Menu Setting Description<br />
operate<br />
Enable Control enable switch<br />
Operate Control operate switch<br />
Output<br />
Pass<br />
Fail<br />
Beep<br />
Stop<br />
Info<br />
Signal tested corresponds with the rule<br />
Signal tested not correspond with the rule<br />
Beep when it satisfies the rule<br />
Stop once satisfying the rule<br />
Control the display status of info frame<br />
Source<br />
Select source CH1, CH2 or Math<br />
Rule<br />
Horizontal<br />
Change the Horizontal tolerance value by turning the M knob<br />
Vertical<br />
Change the Vertical tolerance value by turning the M knob<br />
Create<br />
Use the rule set as testing rule<br />
SaveRule<br />
Number<br />
Save<br />
Load<br />
Choose any one from Rule1~Rule8 as your rule name<br />
Click Save to save the rule<br />
Load some rule as the testing rule<br />
The description of Pass/Fail Menu is shown as the follows:<br />
Pass/Fail test:<br />
Detect whether the input signal is within the limits of the rule, if it exceeds limits of the rule, it is “Fail”; otherwise<br />
it is “Pass”. Also it can output fail or pass signal by built-in, configurable and photoelectric isolation output port.<br />
To run the test, read the following steps:<br />
1. Press Utility button, then H1 button, turn the M knob to choose Pass/fail menu option, Pass/Fail menu<br />
will be displayed in the bottom.<br />
2. Enable switch on: Press H2 button to show Operate menu, then, press F1 button to set Enable as ON.<br />
3. Create rule: Press H4 button to enter Rule setting menu. Press F1 button to choose the source; Press<br />
F2 button, turn the M knob to set Horizontal tolerance; Press F3 button, turn the M knob to set Vertical<br />
tolerance; Press F4 button to create the rule.<br />
4. Set output type: Press H3 button to enter Output option setting. Choose any one or two of the options<br />
“Pass”, “Fail” or “Beep”. “Pass” and “Fail” are mutually exclusive options, which could not be chosen<br />
simultaneously. “Stop” means stop once the condition satisfies your setting.<br />
5. Begin to test: Press H2 button, then F2 button to select “Start”, the test will begin.<br />
6. Save rule: Press H5 button, then F2 button to save the rules, which could be called up at once when<br />
need, press F3 button to call up the rule saved.<br />
-138-