Function Menu Setting Description<br />

operate<br />

Enable Control enable switch<br />

Operate Control operate switch<br />

Output<br />

Pass<br />

Fail<br />

Beep<br />

Stop<br />

Info<br />

Signal tested corresponds with the rule<br />

Signal tested not correspond with the rule<br />

Beep when it satisfies the rule<br />

Stop once satisfying the rule<br />

Control the display status of info frame<br />

Source<br />

Select source CH1, CH2 or Math<br />

Rule<br />

Horizontal<br />

Change the Horizontal tolerance value by turning the M knob<br />

Vertical<br />

Change the Vertical tolerance value by turning the M knob<br />

Create<br />

Use the rule set as testing rule<br />

SaveRule<br />

Number<br />

Save<br />

Load<br />

Choose any one from Rule1~Rule8 as your rule name<br />

Click Save to save the rule<br />

Load some rule as the testing rule<br />

The description of Pass/Fail Menu is shown as the follows:<br />

Pass/Fail test:<br />

Detect whether the input signal is within the limits of the rule, if it exceeds limits of the rule, it is “Fail”; otherwise<br />

it is “Pass”. Also it can output fail or pass signal by built-in, configurable and photoelectric isolation output port.<br />

To run the test, read the following steps:<br />

1. Press Utility button, then H1 button, turn the M knob to choose Pass/fail menu option, Pass/Fail menu<br />

will be displayed in the bottom.<br />

2. Enable switch on: Press H2 button to show Operate menu, then, press F1 button to set Enable as ON.<br />

3. Create rule: Press H4 button to enter Rule setting menu. Press F1 button to choose the source; Press<br />

F2 button, turn the M knob to set Horizontal tolerance; Press F3 button, turn the M knob to set Vertical<br />

tolerance; Press F4 button to create the rule.<br />

4. Set output type: Press H3 button to enter Output option setting. Choose any one or two of the options<br />

“Pass”, “Fail” or “Beep”. “Pass” and “Fail” are mutually exclusive options, which could not be chosen<br />

simultaneously. “Stop” means stop once the condition satisfies your setting.<br />

5. Begin to test: Press H2 button, then F2 button to select “Start”, the test will begin.<br />

6. Save rule: Press H5 button, then F2 button to save the rules, which could be called up at once when<br />

need, press F3 button to call up the rule saved.<br />