32. The Cursor Measurement for FFT model<br />
Press the Cursor button to display the cursor measurement function menu (CURS MEAS) in the screen,<br />
which includes Vamp Measurement and Freq Measurement at the mode of FFT, shown as Fig. 73.<br />
Fig. 73 CURS MEAS Menu<br />
The description of the cursor measurement menu is shown as the following table:<br />
Function Menu Setting Description<br />
Type<br />
OFF<br />
Vamp<br />
Freq<br />
Switch off the cursor measurement.<br />
Display the Vamp measurement cursor and menu.<br />
Display the Freq measurement cursor and menu.<br />
Source Math FFT Display the channel for the cursor measure.<br />
When carrying out the cursor measurement, the position of Cursor 1 can be adjusted with the VERTICAL<br />
POSITION knob of Channel 1, and that of Cursor 2 can be adjusted with the VERTICAL POSITION knob of<br />
Channel 2.<br />
Perform the following operation steps for the Vamp cursor measurement:<br />
1. Press Cursor and recall the Cursor Measure menu.<br />
2. Press the H1 button, the Type menu will display at the right of the screen. Press the F2 button to select<br />
Vamp for Type, with two purple dotted lines displayed along the horizontal direction of the screen<br />
indicating Cursor1 and Cursor2.<br />
3. According to the measured waveform, adjust the positions of Cursor1 and Cursor2 by turning the<br />
VERTICAL POSITION knob of CH1 and CH2. Cursor increment window at the left bottom shows absolute<br />
value of the two cursors amplitude difference and the present position.<br />