Operating Instructions Chapter 7<br />

<strong>CLV63x</strong> <strong>Bar</strong> <strong>Code</strong> <strong>Scanner</strong><br />

Startup and configuration<br />

7.6 Adjusting the bar code scanner<br />

7.6.1 Adjusting the bar code scanner<br />

To completely adjust the bar code scanner, the electrical installation must be complete and<br />

the device must be operated, see chapter 6 Electrical installation, page 49 and chapter 7<br />

Startup and configuration, page 63.<br />

1. Align the bar code scanner in such a way that the angle between the scanning line and<br />

the bar code's lines is almost 90°.<br />

2. To avoid disruptive reflections, rotate the bar code scanner from the plumb line so that<br />

the emitting light meets the bar code at an angle of approx. 105° (line scanner),<br />

chapter 5.3.4 Avoiding surface reflections, page 44.<br />

3. Manually bring objects with bar codes sequentially into the bar code scanner's visual<br />

range. The default setting of the focal position is 285 mm (11.2 in) from the reading<br />

window. Check the reading result using the SOPAS-ET configuration software.<br />

Move objects in different positions (angles) to the reading area and ensure that the<br />

thresholds of the permitted reading angles are not exceeded.<br />

4. Align the bar code scanner in such a way that the good read rate is between 70 and<br />

100%.<br />

5. Tighten the screws on the bar code scanner.<br />

The bar code scanner is aligned with the bar code.<br />

8011970/S345/2008-04-16 © SICK AG · Division Auto Ident · Germany · All rights reserved 69