Vectra Genisys Service Manual - DJO Global

Vectra Genisys Service Manual - DJO Global

Vectra Genisys Service Manual - DJO Global

SHOW MORE
SHOW LESS
  • No tags were found...

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

5- TROUBLESHOOTING<strong>Vectra</strong> <strong>Genisys</strong>®/Intelect Legend XT®/ Intelect® Vet Therapy Systems5.1 THERAPY SYSTEM ERROR MESSAGES (CONTINUED)CodeNumberTypeMessageCRITICAL ERRORS DEMANDING TECHNICAL SERVICE (300-399)Probable CausePossible Remedies300 Critical Error Stim Board not found on Power up. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.301 Critical Error Stim Board failed Power up self test. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.302 Critical Error Stim Board Reset occurred. But, neither the main uPnor channel uP reset bit was set.A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.303 Critical Error Main uP on Stim Board reset occurred. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.304 Critical Error Channel A uP on Stim Board reset occurred. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.305 Critical Error Channel B uP on Stim Board reset occurred. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.306 Critical Error Error writing to Stim Board. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.307 Critical Error Error writing from Stim Board. A. On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and ControlBoard.B. On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module RibbonCable is seated in the Therapy System and on the Module.C. If problem persists, replace appropriate Stim PC Board.D. Replace Control Board.35

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!