6- REMOVAL/REPLACEMENT<strong>Vectra</strong> <strong>Genisys</strong>®/Intelect Legend XT®/ Intelect® Vet Therapy Systems6.3 THERAPY SYSTEM- SEPARATING TOP &BOTTOM (CONTINUED)DISCONNECT THE SYSTEM FROM THE POWERSOURCE (OUTLET OR REMOVE BATTERY MODULE IFINSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,INSTALLATION, REMOVAL, OR REPLACEMENTPROCEDURES TO PREVENT ELECTRICAL SHOCK ANDPOSSIBLE DAMAGE TO SYSTEM.6. Lay System Top on the edge and disconnectthe remaining harnesses from the ControlBoard. See Figure 6.21.C. Replacing Top to BottomReplace System Top by reversing the steps 1-5on page 66.NOTE:Do not over tighten the screws. Over tighteningwill damage the threads of the brass inserts.REMOVEHARNESSESFIGURE 6.2167