5- TROUBLESHOOTING5.9 PREMODULATED MODE TESTSet up System per 5.6 parts A and B prior toperforming test.A. Premodulated Mode Test Procedures1. Set Scope; Time- 2.50 mS, Channel- 20 V, andTrigger- DC.2. Press Electrotherapy Button.3. Press IFC Premodulated (2p) Button.4. Rotate Intensity Control clockwise until 50is displayed.5. Press “Start”.6. Compare waveform form on scope toFigure 5.7.7. Press Stop. Then press Home Button andmove scope probes to Channel 2 and repeatsteps 2 through 6. Repeat test on channels3 and 4 on 4 Channel Systems.B. Premodulated Mode Test Results1. Waveform is the same between scope andFigure 5.7.Unit passed test.2. No waveform or considerably differentwaveform.Unit failed test. Replace appropriate StimBoard.<strong>Vectra</strong> <strong>Genisys</strong>®/Intelect Legend XT®/ Intelect® Vet Therapy SystemsFIGURE 5.743