5- TROUBLESHOOTING5.12 HIGH VOLTAGE PULSED CURRENT(HVPC) MODE TESTSet up Unit per 5.6 parts A and B prior to performingtests.A. High Voltage Pulsed Current (HVPC) ModeTest Procedures1.2.3.4.5.6.7.8.9.10.11.Set Scope; Time- 25 μS, Channel- 50 V, andTrigger- DC.Press Electrotherapy Button.Press High Volt Button.Rotate Therapy Intensity Control clockwiseuntil 250 Volts is displayed.Press Start Button.Compare waveform on scope to Figure5.12.Press Edit Button.Press Polarity Button until Positive isdisplayed in Polarity icon.Compare waveform form on scope toFigure 5.13.Press Display Button until Peak Current isdisplayed in the Display icon.The numbers displayed for amplitudemust not exceed 1.5 Amps. See Figure5.14.Press “Stop”.12.B. High Voltage Pulsed Current (HVPC) ModeTest Results1. Waveforms on scope the same as Figures5.12 and 5.13 and Amps do notexceed 1.5.Unit passed test.2. No waveform or considerably differentwaveforms.Unit failed test. Replace appropriate StimBoard.3. Amps exceed 1.5.Unit failed test. Replace appropriate StimBoard.AMPS MUST NOTEXCEED 1.5<strong>Vectra</strong> <strong>Genisys</strong>®/Intelect Legend XT®/ Intelect® Vet Therapy SystemsPROPER NEGATIVE HIGH VOLT WAVEFORMFIGURE 5.12PROPER POSITIVE HIGH VOLT WAVEFORMFIGURE 5.13FIGURE 5.1446