INSTRUMENT OPERATION MODESTo access standard name menu:1. Use the Tab keys $@ to highlight Std Name.Edit Standard : 1Std Entry:MeasuredStd Name :Tolerances...Shade Sort Opts...Std Lock :UnlockedDelete This Std...2. Press the Enter key # to access the Edit StandardName menu.NOTE: To quickly remove a name, highlighting CLEAR and press theEnter key #.↓↑Edit Standard NameCLEARoName entry fieldSave & Exit3. Use the Tab keys $@ to highlight the name entry field.4. Use the Tab keys $@ to choose the desired characterlocation (arrows above and below designate the selection).Press the Enter key # to access the alphanumeric editor.5. Press the Enter key # again to quickly page throughgroups of letters, symbols, and numbers.6. Use the Tab keys $@ to highlight the desired character andpress the Enter # key to exit editor.7. Continue with additional character edits.8. When editing is completed, use the Tab keys $@ tohighlight Save & Exit and press the Enter # key.5.1.4 Setting Tolerance LimitsThe tolerance limit is the maximum allowable difference fromthe standard color values that is considered acceptable. Thetolerances are used to test your sample’s acceptability bydisplaying a pass or fail signal, based on the entered limits. Theplus and minus limits can be set equal or individually forL*a*b*, L*C*h°, and L*u*v*. Pass/fail indication appears inQA and Strength modes when set.5-5