08.01.2014 Views

Gisborne Hospital Report - Health and Disability Commissioner

Gisborne Hospital Report - Health and Disability Commissioner

Gisborne Hospital Report - Health and Disability Commissioner

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Gisborne</strong> <strong>Hospital</strong> 1999 – 2000<br />

PSA Testing Procedures<br />

2.17 External quality assurance refers to programmes provided by an independent<br />

external source, such as the Australasian Quality Assurance Programme<br />

(AQAP) provided by the Royal College of Pathologists of Australasia. AQAP<br />

is used by the majority of laboratories in New Zeal<strong>and</strong> as their external QA<br />

programme.<br />

2.18 The programme organisers send out samples at regular intervals (monthly for<br />

PSA) to a large number of subscribing laboratories. The samples have target<br />

values for many different analytes, but these target values are unknown to the<br />

testing laboratory. After analysing the external control sample, the laboratory<br />

submits their results to the organisers. A report is then returned to the<br />

laboratory, which compares their submitted results with those submitted from<br />

all other laboratories, <strong>and</strong> they are also compared with the target values.<br />

2.19 The external quality assurance programme thus acts as a final safety net for all<br />

other quality control procedures. If a result does not compare closely with the<br />

results obtained by the majority of other laboratories using the same<br />

instrument <strong>and</strong> same reagents, then it is clear that there is a quality problem.<br />

The presentation of data in the external quality control programme reports<br />

illustrate the type of problem a laboratory may be having.<br />

Calibrators<br />

2.20 A calibrator is a sample containing an accurately measured <strong>and</strong> known amount<br />

of a particular analyte. When an instrument is calibrated, at least two<br />

calibrators are measured, one with a low value <strong>and</strong> one with a high value. The<br />

instrument uses the readings it gets from these calibrators to draw a line (or<br />

curve if more than two calibrators), <strong>and</strong> this curve is then used to relate the<br />

read-out of the instrument to a concentration of the analyte. In effect,<br />

instrument readings on subsequent patient <strong>and</strong> internal control specimens are<br />

compared with the calibration curve to calculate the concentration of the<br />

analyte.<br />

2.21 If the wrong calibrators are used, as they were at <strong>Gisborne</strong> <strong>Hospital</strong> for PSA,<br />

the calibration curve is incorrect <strong>and</strong> the concentrations of PSA calculated in<br />

internal control <strong>and</strong> patient specimens are also incorrect. The error with the<br />

internal control results was recognised, but by altering the control values, the<br />

problem was dealt with inappropriately - in effect, ignored - <strong>and</strong> the errors<br />

continued with patient specimens.<br />

2.22 Because many factors are involved in maintaining quality, a system of records<br />

<strong>and</strong> documentation is needed for periodic review <strong>and</strong> evaluation by the<br />

laboratory, but also by agencies such as IANZ. These quality control records<br />

must be maintained for a period of time (often a minimum of two years) to<br />

document that testing has occurred. These records should include:<br />

• Routine maintenance<br />

• Reagent lot numbers in use <strong>and</strong> expiration dates<br />

139

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!