54<br />

After polarization, the beam was amplitude modulated with a st<strong>and</strong>ard optical<br />

chopper (1 kHz) <strong>and</strong> a sequence <strong>of</strong> optics were used to direct the beam to a beam<br />

splitter which transmitted ,,-,90% <strong>of</strong> the fundamental light to the sample. The remaining<br />

10% <strong>of</strong> the Ti:Ah03 pulse was reflected <strong>and</strong> propagated along an identical<br />

optical line containing a quartz wedge. This line was used to normalize the nonlinear<br />

optical signal, which will be described in greater detail in a following paragraphs. The<br />

sample was placed at the focal point <strong>of</strong> an fj 4.0 achromatic lens which focused the<br />

4 mm beam to 100 J.Lm spot. The achromatic lens has a Rayleigh range <strong>of</strong> 4 mm which<br />

ensured that the spot size <strong>of</strong> the beam did not vary significantly as the wavelength<br />

<strong>of</strong> the fundamental beam was scanned. Any second-harmonic signal generated along<br />

the beam path prior to the sample was removed with a low pass optical filter.<br />

Interaction <strong>of</strong> the ultrafast Ti:Ah03 laser pulse with the sample or quartz produced<br />

a second-harmonic generation pulse with angular frequency 2w. The SHG<br />

pulse was separated from the fundamental pulse by high pass filters <strong>and</strong> a Jarrel Ash<br />

:\Ionospec 18 monochrometer with entrance <strong>and</strong> exit slit widths <strong>of</strong> 2 mm. This slit<br />

configuration has a top hat transmission function with a full width wavelength b<strong>and</strong>width<br />

<strong>of</strong> ~,,\<br />

= 10 nm. The monochrometer grating was blazed at 500 nm to maximize<br />

sensitivity to near IR frequency doubled light. In addition, the SHG signal was pol::trized<br />

to probe various second-order susceptibility elements. The SHG signal was<br />

detected by a Hamamatsu R943-02 photomultiplier tube. The R943-02 is a high<br />

Reproduced with permission <strong>of</strong> the copyright owner. Further reproduction prohibited without permission.