Renamic - BIOTRONIK USA - News
Renamic - BIOTRONIK USA - News
Renamic - BIOTRONIK USA - News
- No tags were found...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
49 <strong>Renamic</strong> PSAFigure 20: Warning message for ventricular burst pacingReportPress the OK button on the message screen to continue.Use the slide bar to increase the pacing rate in theprogrammed chamber. Releasing the button returns the rateto 80 bpm.NOTE: Emergency medical equipment should always beavailable when burst pacing is performed.The report page, shown in Figure 21, presents a summaryof saved data from testing. The data is stored in theprogrammer, and this page may be printed, as well. Ifmultiple testing occurs, the last data point tested will bestored for that session.