53 <strong>Renamic</strong> PSAFigure 24: Impedance testThresholdTo start the threshold test, first select the chamber beingmeasured under the Save measurement section. An exampleis shown in the blue box in Figure 25. Press the amplitudebutton (black box) for the chamber being tested, in this casethe RV channel. Decrease the output by using the slidebaror – button until capture is lost. Increase the output toregain threshold and press the Threshold button under theSave measurement section to save the data to the Reportscreen. A message stating “Values saved” will appear on themessage bar, as shown with the red box in Figure 25.Increase the output if the patient is pacemaker dependent toregain capture or decrease the rate to end the test.