digiset 5<br />

4.2.2 Check illumination<br />

After the working distance is checked, the homogeneity and brightness of the illumination<br />

must be set. For this, you must not have anything in the camera’s visual field.<br />

To check the illumination the greyscales are checked in the camera’s horizontal and<br />

vertical alignment on both edges of the image as well as the middle of the image and<br />

the results shown on 6 lines.<br />

The illumination reaches its optimum setting if these lines are as horizontal as possible<br />

and within the horizontal tolerance range marked by 2 purple lines.<br />

You access the operating mode for setting the illumination using the following menu<br />

steps:<br />

Main menu Sub-menu 1 Sub-menu 2<br />

Setup � Imaging � Illumination<br />

The TCAM2 displays the following image with 6 grey scales (illumination lines):<br />

Tolerance field<br />

Procedure:<br />

1. Select the illumination check with the "Illumination" operating mode. The<br />

horizontal tolerance field appears with the illumination lines. "Illumination"<br />

appears in the status window.<br />

2. If not all 6 greyscale lines on the LCD are inside the tolerance field, you<br />

have to adjust the light source. Remove the cover of the light source as<br />

shown on the picture below.<br />