digiset 5<br />
8.2.2 Maximum<br />
The measuring lines are created on the contour of the object to be measured. Both<br />
adjustment bars are hidden. The offset of the measuring lines and the fixed cross<br />
hairs in the X and Z direction is taken into account with the current position.<br />
You access this operating mode using the following menu steps:<br />
Main menu Sub-menu 1<br />
Imaging � Maximum<br />
After selection the system analyses the object contour and automatically suggests a<br />
measuring procedure for carrying out the measurements immediately. The measuring<br />
procedure used is displayed in the status window at the same time. If an alternative<br />
measuring method is required it is possible to enter the measuring procedure manually<br />
by pressing the button.<br />
Procedure:<br />
�<br />
1. Move the object into the camera’s visual range.<br />
2. If the relevant contour of the object is not clear, select the "Range" operating<br />
mode and, using the mouse, make it the required size and drag it round the<br />
relevant edge.<br />
3. Select the operating mode as described above.<br />
4. Turning the object to be measured on its support sets the greatest deviation.<br />
5. While the object is being turned the test LED turns green as soon as the<br />
measurement is correct, i.e. the greatest deviation has been determined.<br />
You must ensure here that the turning speed is not too great. When the<br />
measurement has been made correctly the offset values are saved and<br />
transferred through the serial port.<br />
6. If the object is set to a specific size, the object must be moved again to the<br />
greatest deviation position so that the arrow of the adjustment bar corresponding<br />
to the deviation direction is in the green area again.<br />
7. Then the unit switches to the "Preset" measuring method.<br />
<strong>Urma</strong> <strong>AG</strong>, Switzerland www.urma.ch page 43