digiset 5<br />

8.3.2.5 "Line 1 - Point X" measurement (L1 - PX)<br />

Linie 1<br />

Punkt X<br />

The vertical axis is touched at one point and shown as a red<br />

measuring line. Line 1 runs through the cross hairs origin point.<br />

The contour is aligned to its optimum level when the pointers on<br />

both adjustment bars are exactly in the middle.<br />

8.3.2.6 "Line 2 - Point X" measurement (L2 - PX)<br />

Linie 2 Punkt X<br />

The maximum X deviation and line 2 are determined. The results<br />

then show the intersection point.<br />

8.3.2.7 "Line 1 - Point Z" measurement (L1 - PZ)<br />

Linie 1<br />

The maximum Z deviation and line 1 are determined. The results<br />

then show the intersection point.<br />

8.3.2.8 "Line 2 - Point Z" measurement (L2 - PZ)<br />

Punkt Z<br />

Punkt Z<br />

Linie 2<br />

The horizontal axis is touched at one point and shown as a red<br />

measuring line. Line 2 runs through the cross hairs origin point.<br />

The contour is aligned to its optimum level when the pointers on<br />

both adjustment bars are exactly in the middle.<br />