6 Service and Maintenance<br />
<strong>360</strong> IG TURN ON<br />
BEGIN<br />
IG1 OR IG2<br />
SWITCH ON<br />
SELECT CORRECT<br />
RANGE<br />
SELECT CORRECT<br />
CATHODE<br />
REPLACE CABLE<br />
NO NO YES<br />
DOES<br />
IG SWITCH<br />
LAMP STAY<br />
ON?<br />
NO<br />
CHECK PRESSURE<br />
RANGE SWITCH<br />
IS<br />
PRESSURE<br />
RANGE SWITCH<br />
SET<br />
CORRECTLY?<br />
YES<br />
CHECK CATHODE<br />
SWITCH<br />
IS<br />
CATHODE<br />
SWITCH<br />
SET<br />
CORRECTLY?<br />
YES<br />
CHECK<br />
CONTROLLER<br />
INTERCONNECT<br />
CABLE<br />
IS<br />
CONTR.<br />
INTERCONNECT<br />
CABLE<br />
DAMAGED?<br />
NO<br />
YES<br />
CHECK<br />
STABIL-ION<br />
CABLE<br />
IS<br />
STABIL-ION<br />
CABLE<br />
DAMAGED?<br />
NO<br />
CHECK<br />
STABIL-ION<br />
GAUGE<br />
IS<br />
STABIL-ION<br />
GAUGE<br />
DAMAGED?<br />
NO<br />
CALL FACTORY<br />
YES<br />
YES<br />
<strong>360</strong> IG TURN ON<br />
FINISH<br />
REPLACE CABLE<br />
REPLACE<br />
STABIL-ION GAUGE<br />
Figure 6-8<br />
<strong>Stabil</strong>-<strong>Ion</strong> Gauge Turn On.<br />
IG PRESSURE<br />
DISPLAY BEGIN<br />
READ PRESSURE<br />
DISPLAY<br />
SELECT CORRECT<br />
UNITS<br />
SELECT CORRECT<br />
SCALE FACTOR<br />
SET CORRECT<br />
RANGE<br />
NO NO NO<br />
IS<br />
PRESSURE<br />
DISPLAY AS<br />
EXPECTED?<br />
NO<br />
CHECK DISPLAY<br />
UNITS SWITCH<br />
IS<br />
SWITCH<br />
SET<br />
CORRECTLY?<br />
YES<br />
CHECK SCALE<br />
FACTOR SWITCH<br />
IS<br />
SCALE<br />
FACTOR<br />
SET<br />
CORRECTLY?<br />
YES<br />
CHECK<br />
PRESSURE<br />
RANGE SWITCH<br />
IS<br />
RANGE<br />
SWITCH<br />
SET<br />
CORRECTLY?<br />
YES<br />
YES<br />
CHECK<br />
STABIL-ION<br />
GAUGE<br />
IS<br />
STABIL-ION<br />
CABLE<br />
DIRTY?<br />
NO<br />
CHECK<br />
STABIL-ION GAUGE<br />
COLLECTOR CABLE<br />
IS<br />
CABLE<br />
DAMAGED?<br />
NO<br />
CALL FACTORY<br />
YES<br />
YES<br />
IG PRESSURE<br />
DISPLAY FINISH<br />
DEGAS<br />
STABIL-ION GAUGE<br />
REPLACE CABLE<br />
Figure 6-9<br />
<strong>Stabil</strong>-<strong>Ion</strong> Gauge Display.<br />
6-12<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001