3 Installation<br />
3.7.3 Using the Remote Outputs<br />
The relay contact outputs are accessible on the 15 pin remote output connector. These contacts<br />
provide status indication for IG1 and IG2 on or off, degas on or off, and fault detection. Figure 3-13<br />
shows the pin designations. See Section 6.3 on page 6-4 for a description of fault relay indications.<br />
Figure 3-13<br />
Remote Output Connector (Use DA-155 Mating Connectors, Supplied).<br />
3.8 Connecting the RS-232 Computer Interface<br />
Instructions for setting up this interface are in Section 2.1.5 on page 2-9.<br />
This factory or field installed capability produces the signals shown in Table 3-4.<br />
A mating DB-25S connector is supplied in the hardware kit. Use shielded cable to minimize<br />
electromagnetic radiation or susceptibility. Ground the shield to the metal connector shell or to<br />
Pin 1. Do not connect the shield to Pin 7.<br />
REMOTE<br />
OUTPUT<br />
9 1<br />
15 8<br />
GND<br />
POWER<br />
SUPPLY<br />
1 13<br />
12<br />
25<br />
PROCESS CONTROL<br />
C<br />
X<br />
A<br />
W<br />
ANALOG<br />
OUTPUT A<br />
1 9<br />
8<br />
15<br />
C<br />
G<br />
G<br />
A<br />
U<br />
G<br />
E<br />
ANALOG<br />
OUTPUT B<br />
ELECTROMETER<br />
IG1 COL.<br />
ANA.<br />
OUT<br />
IG2<br />
1 19<br />
R<br />
E<br />
M<br />
O<br />
T<br />
E<br />
9 26<br />
COL.<br />
MIN.ADJ.<br />
RS-232<br />
13 25<br />
1 14<br />
DB-25P connector<br />
Figure 3-14<br />
Control Unit Rear Panel Showing RS-232 Connector.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
3-13