2 Initial Setup Procedures<br />
2.1.5.1.2 Talk-Only Mode<br />
The pressure data from all three displays will be output in a single message string, separated by<br />
commas, approximately every 5 seconds. If the Convectron Module is not present, only the ion<br />
gauge pressure will be output.<br />
Table 2-7<br />
Talk-Only Mode.<br />
S1<br />
Off<br />
On<br />
Mode<br />
Talk-only<br />
Command-response*<br />
1<br />
1<br />
OFF<br />
ON<br />
*factory setting<br />
\<br />
Figure 2-16<br />
Partial Top View of RS-232<br />
Module.<br />
2.1.5.2 Handshaking<br />
Table 2-8 Inputs to <strong>360</strong>.<br />
Line Pin Description<br />
Internal Switch<br />
Function<br />
Factory<br />
Setting<br />
CTS<br />
&<br />
DSR<br />
5<br />
6<br />
CLEAR to SEND<br />
and DATA SET<br />
READY: When<br />
used, both must<br />
be TRUE in order<br />
for <strong>360</strong> to send<br />
the next byte in its<br />
message or data.<br />
CTS=1 and DSR=1:<br />
When ON, forces the<br />
functions TRUE and<br />
thus assumes host is<br />
always ready to<br />
receive.<br />
Both ON<br />
DCD 8 DATA CARRIER<br />
DETECT: Must be<br />
TRUE at the time<br />
each character is<br />
received or that<br />
character will be<br />
ignored by <strong>360</strong>.<br />
DCD=1: When ON,<br />
forces DCD function<br />
TRUE so <strong>360</strong> will<br />
receive all characters<br />
sent to it (as long as<br />
RTS is in<br />
de-asserted state).<br />
ON<br />
X<br />
X<br />
OFF<br />
ON<br />
Figure 2-17 Top View of RS-232 Module.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
2-11