2 Initial Setup Procedures<br />
2.1.6.1 Selecting the Interface Bus Address<br />
Refer to Figure 2-19 and locate switches 1 through 5. Each instrument on the IEEE 488 bus has an<br />
address from 0 to 30. The address for this product is set at the factory to “8”. The user must ensure<br />
that every device on his bus has a unique address. To select an address, switches 1 through 5 are<br />
adjusted according to binary weights:<br />
Table 2-10 Interface Bus Address Values When OFF.<br />
Switch Number 5 4 3 2 1<br />
Value when OFF 16 8 4 2 1<br />
Examples:<br />
Table 2-11 Example Switch Settings.<br />
Switch Number 5 4 3 2 1 Address<br />
Setting<br />
On<br />
On<br />
Off<br />
* Factory default setting<br />
Off<br />
On<br />
Off<br />
On<br />
Off<br />
Off<br />
On<br />
Off<br />
On<br />
On<br />
Off<br />
On<br />
8*<br />
7<br />
28<br />
X<br />
X<br />
OFF<br />
ON<br />
2.1.6.2 Talk-Only Mode<br />
Figure 2-19<br />
Switch 8, if set to off during power up, places the interface in<br />
talk-only mode. The pressure data from all three displays will<br />
be output in a single message string, separated by commas,<br />
approximately every 5 seconds. If no Dual Convectron Gauge<br />
module is present, only the ion gauge pressure will be output.<br />
The factory default is talk-only disabled (switch 8 on).<br />
8<br />
8<br />
Top View of IEEE 488 Module.<br />
OFF<br />
ON<br />
2.1.6.3 SRQ Mode<br />
Figure 2-20<br />
Switch 7, if set to off during power up, tells the interface to<br />
generate an SRQ interrupt and expect a serial poll before every<br />
message output. The SRQ bit will be released when there is a<br />
Serial Poll or when the message is read. The factory default is<br />
SRQ mode disabled (switch 7 on). The serial poll response byte<br />
will contain a 1 in bit 4 (this bit is also the MAV (message<br />
available) bit). Bit 6 will be set if an error has occurred.<br />
In this case the pending message will be an error message.<br />
Switch 6 is not used.<br />
Figure 2-21<br />
Top View of IEEE 488 Module.<br />
OFF<br />
7<br />
ON<br />
7<br />
Top View of IEEE 488 Module.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
2-13