2 Initial Setup Procedures<br />

2.1.6.1 Selecting the Interface Bus Address<br />

Refer to Figure 2-19 and locate switches 1 through 5. Each instrument on the IEEE 488 bus has an<br />

address from 0 to 30. The address for this product is set at the factory to “8”. The user must ensure<br />

that every device on his bus has a unique address. To select an address, switches 1 through 5 are<br />

adjusted according to binary weights:<br />

Table 2-10 Interface Bus Address Values When OFF.<br />

Switch Number 5 4 3 2 1<br />

Value when OFF 16 8 4 2 1<br />

Examples:<br />

Table 2-11 Example Switch Settings.<br />

Switch Number 5 4 3 2 1 Address<br />

Setting<br />

On<br />

On<br />

Off<br />

* Factory default setting<br />

Off<br />

On<br />

Off<br />

On<br />

Off<br />

Off<br />

On<br />

Off<br />

On<br />

On<br />

Off<br />

On<br />

8*<br />

7<br />

28<br />

X<br />

X<br />

OFF<br />

ON<br />

2.1.6.2 Talk-Only Mode<br />

Figure 2-19<br />

Switch 8, if set to off during power up, places the interface in<br />

talk-only mode. The pressure data from all three displays will<br />

be output in a single message string, separated by commas,<br />

approximately every 5 seconds. If no Dual Convectron Gauge<br />

module is present, only the ion gauge pressure will be output.<br />

The factory default is talk-only disabled (switch 8 on).<br />

8<br />

8<br />

Top View of IEEE 488 Module.<br />

OFF<br />

ON<br />

2.1.6.3 SRQ Mode<br />

Figure 2-20<br />

Switch 7, if set to off during power up, tells the interface to<br />

generate an SRQ interrupt and expect a serial poll before every<br />

message output. The SRQ bit will be released when there is a<br />

Serial Poll or when the message is read. The factory default is<br />

SRQ mode disabled (switch 7 on). The serial poll response byte<br />

will contain a 1 in bit 4 (this bit is also the MAV (message<br />

available) bit). Bit 6 will be set if an error has occurred.<br />

In this case the pending message will be an error message.<br />

Switch 6 is not used.<br />

Figure 2-21<br />

Top View of IEEE 488 Module.<br />

OFF<br />

7<br />

ON<br />

7<br />

Top View of IEEE 488 Module.<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />

2-13