1 System Components<br />
10<br />
9<br />
8<br />
POWER 50-60 HZ 220 WATTS<br />
IG2<br />
IG1<br />
POWER<br />
FUSE<br />
DEGAS<br />
FUSE<br />
!<br />
ALL 3 SWITCHES MUST BE SET TO THE SAME AC LINE VOLTAGE<br />
!<br />
FOR AC LINE VOLTAGE BETWEEN 90V TO 130V SET ALL SWITCHES TO 115V,<br />
USE 3A, F 250V POWER FUSE, AND USE 2A, F, 250V DEGAS FUSE<br />
FOR AC LINE VOLTAGE BETWEEN 180V TO 240V SET ALL SWITCHES TO 230V,<br />
USE 1.6A, F, 250V POWER FUSE, AND USE 1A, F, 250V DEGAS FUSE<br />
CONTROL UNIT<br />
1 2 3 4 5 6 7<br />
Figure 1-7<br />
<strong>Stabil</strong>-<strong>Ion</strong> Power Supply.*<br />
1. Line power fuse 6. Line voltage selector switch<br />
2. Grounding lug (connect to earth ground with<br />
#12 AWG conductor)<br />
7. Connector for interconnect cable to Control Unit<br />
3. Degas power fuse 8. Connector for <strong>Stabil</strong>-<strong>Ion</strong> Gauge 1<br />
4. Line voltage selector switch 9. Connector for <strong>Stabil</strong>-<strong>Ion</strong> Gauge 2<br />
5. Line voltage selector switch 10. Connector for power cable<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
1-5