3 Installation<br />
1. Using Table 3-1 and circuit schematics you have prepared, make up a cable to connect the<br />
various system components which are to be controlled. Unambiguous labeling of each lead<br />
will help prevent costly mistakes.<br />
RS-232<br />
ELECTROMETER<br />
CONVECTRON<br />
IG-1 AUTO<br />
PROCESS CONTROL<br />
Off SP On<br />
Data Port<br />
IG2 Scale<br />
IG1<br />
Factor<br />
Adjust<br />
Pressure<br />
Range<br />
L HL H<br />
Cathode<br />
1 21 2<br />
IG1 IG2<br />
OFF SET<br />
ATM<br />
VAC<br />
IG-2 AUTO<br />
OFF SET<br />
ATM<br />
VAC<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
3<br />
▲<br />
Set<br />
▲<br />
C<br />
H<br />
A<br />
N<br />
N<br />
E<br />
L<br />
IG1<br />
O<br />
Degas<br />
O<br />
IG2<br />
O<br />
On<br />
O<br />
Remote<br />
Figure 3-9<br />
2. Ensure that the Process Control channel override switches are all set to off.<br />
3. Connect the component end of the cable to the system component to be controlled.<br />
4. Plug the connector into the back of the Control Unit.<br />
5. Refer to Section 4.11 on page 4-26 for instructions for setting setpoints.<br />
3.7 Connecting <strong>Stabil</strong>-<strong>Ion</strong> Remote Control<br />
Front Panel of Control Unit (with Door Open) Showing Locations of Override<br />
Switches.<br />
This factory installed option permits the following functions to be controlled from a remote<br />
location:<br />
■ <strong>Ion</strong>ization gauge IG1/IG2 on/off ■ Pressure range selection<br />
■ Degas on/off ■ <strong>Ion</strong>ization gauge lockout<br />
■ Cathode select 1 or 2 ■ Front panel switch disable<br />
In addition, the following functions are outputted in the form of relay contact closures which can<br />
signal the status of the instrument to a remote location, or control external devices.<br />
■ IG-1 on/off ■ Degas on/off<br />
■ IG-2 on/off ■ Circuit fault detection<br />
The optically isolated control inputs are on a 26 pin remote input connector. The relay contact<br />
outputs are accessible on the 15 pin remote output connector.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
3-9