Chapter 5<br />

5 Operation of a <strong>Stabil</strong>-<strong>Ion</strong> System<br />

The instructions in this chapter assume the instructions for Setup, Installation, and Preparing For<br />

Operation have been completed. See Chapters 2, 3, and 4.<br />

1. Press the power ON switch on the<br />

Control Unit (see Figure 5-1).<br />

2. If Convectron Gauge capability<br />

is installed, Convectron Gauge<br />

pressures will be displayed on lines<br />

A and B of the display whenever the<br />

power switch is on.<br />

IG 8.56 -7<br />

A<br />

B<br />

9.43 -3<br />

5.67 -2<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

IG1<br />

O<br />

IG2<br />

O<br />

DEGAS ON<br />

O O<br />

Figure 5-1<br />

Control Unit Front Panel Showing<br />

Power ON Button.<br />

Table 5-1<br />

Convectron Gauge Display Formats.<br />

Units Display Format Pressure Example<br />

torr scientific < 1 Torr 3.23–3 Torr<br />

floating point > 1 Torr 7.15 Torr<br />

mbar scientific < 1 mbar 5.18–2 mbar<br />

floating point > 1 mbar 88.6 mbar<br />

scientific > 999 mbar 1.23+3 mbar<br />

pascal scientific < 1 Pa 7.23–1 Pa<br />

floating point > 1 Pa 78.3 Pa<br />

scientific > 999 Pa 1.25+4 Pa<br />

Special Considerations for Convectron Gauge Use Below 10 -3 Torr<br />

During a fast pumpdown from atmosphere, thermal effects will prevent the Convectron Gauge<br />

from tracking pressure accurately below 10 -3 Torr. After about 15 minutes, indications in the<br />

10 -4 range will be valid and response will be rapid. Zero adjustment at vacuum may be<br />

performed at this time (or sooner if readings in the 10 -4 range are not needed).<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />

5-1