3 Installation<br />

3.11 Connecting the Power Cables<br />

1. After completing the set up, connect the interconnect cable to the Control Unit and to the<br />

Power Supply. A 0.6 m (2 ft) [or a 3 m (10 ft) for remote mounting] interconnect cable is<br />

supplied with your Controller and has identical 25 pin connectors at both ends. Secure the<br />

ends at the Power Supply connector and the Control Unit connector using the two captive<br />

screws on each connector.<br />

Figure 3-18<br />

Power Supply and Control Unit Rear Views Showing Power Cable Connections.<br />

Compliance with the usual warning to connect the power cable only to a<br />

properly grounded outlet is necessary but not sufficient for safe operation of<br />

a vacuum system with this or any similar high voltage producing product.<br />

Grounding this product does not and cannot guarantee that other<br />

components of the vacuum system are all maintained at earth ground.<br />

It is the installer's responsibility to ensure that the automatic signals<br />

provided by the product are always used in a safe manner. Carefully check<br />

the system using manual operation and the setpoint programming before<br />

switching to automatic operation.<br />

Where an equipment malfunction could cause a hazardous situation, always<br />

provide for fail-safe operation. As an example, in an automatic backfill<br />

operation where a malfunction might cause high internal pressures, provide<br />

an appropriate pressure relief device.<br />

Installation and wiring must be in compliance with your local building and<br />

electrical codes. CONNECT THE POWER CORD ONLY TO A PROPERLY<br />

GROUNDED OUTLET.<br />

Primary Mains disconnect is provided by the A.C. Power Cord. Install unit so<br />

there is access to the A.C. Power Cord.<br />

2. If you have not already done so, verify that the line voltage switches are set properly. See<br />

Section 2.2 on page 2-15.<br />

3. Connect the power cable to a grounded outlet.<br />

4. Assuming you have completed the set up and installation according to the foregoing<br />

instructions, the <strong>Stabil</strong>-<strong>Ion</strong> System should now be ready to turn on and operate. If you are<br />

using the Process Control Option, you must prepare for process control operation before<br />

turning on the <strong>360</strong> in automatic mode. See Section 4.11 on page 4-26. Attempting automatic<br />

operation with process control before the controls are properly set can cause costly damage<br />

and even be dangerous.<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />

3-19