4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for Operation<br />
4.2.1 Pressure Indication During Degas<br />
An approximate indication of pressure within the <strong>Stabil</strong>-<strong>Ion</strong> Gauge is provided on the digital display<br />
during degas. Observing this pressure indication rise, peak out and then fall is an excellent means<br />
of determining the optimum time of degas. The indication may eventually be unrealistically low due<br />
to a reverse x-ray current effect or temporary gauge pumping. Degassing purposely changes surface<br />
conditions in the gauge. After degassing, permit sufficient time for steady state conditions to be<br />
established before attempting to make measurements.<br />
4.2.2 Degas Power Timer<br />
The degas power timer is factory preset for 10 minutes. See Section 2.1.7 on page 2-14 to change<br />
this time.<br />
4.2.3 Degas Power Limiter<br />
For convenience and improved safety, degas power is automatically reduced as gas is liberated to<br />
prevent excessive pressure rise and help reduce the possibility of a gas discharge forming. As the<br />
electrodes clean up, power is increased to a maximum of 40 W. This feature limits the maximum<br />
pressure indication to about 5 x 10 -5 Torr during degas.<br />
4.2.4 Gauge Bakeout<br />
Applying an external heat source to the vacuum system and to the gauge envelope can speed the<br />
removal of contamination from the interior surfaces. Bakeout temperatures between 150 °C and<br />
300 °C are typically adequate. Do not exceed 450 °C at the <strong>Stabil</strong>-<strong>Ion</strong> Gauge envelope. Do not<br />
exceed 150 °C at the cable and connector.<br />
4.3 Alternate Cathode Operation<br />
The cathode coating, which is utilized in the <strong>Stabil</strong>-<strong>Ion</strong> Gauge to provide electron emission at<br />
relatively low cathode temperatures, has a finite operating life—typically two to three years at low<br />
pressure. <strong>Ion</strong> bombardment at higher pressures tends to erode the coating faster than at lower<br />
pressures. <strong>Stabil</strong>-<strong>Ion</strong> Gauges have a second independent cathode for use when the first cathode<br />
finally tires and does not emit sufficiently. Note, the Controller will automatically turn the IG off if<br />
the cathode has insufficient emission either due to contamination of the cathode or long use.<br />
RS-232<br />
ELECTROMETER<br />
CONVECTRON<br />
IG-1 AUTO<br />
PROCESS CONTROL<br />
Off SP On<br />
Data Port<br />
IG2 Scale<br />
IG1<br />
Factor<br />
Adjust<br />
Pressure<br />
Range<br />
L HL H<br />
Cathode<br />
1 21 2<br />
IG1 IG2<br />
OFF SET<br />
ATM<br />
VAC<br />
IG-2 AUTO<br />
OFF SET<br />
ATM<br />
VAC<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
3<br />
▲<br />
Set<br />
▲<br />
C<br />
H<br />
A<br />
N<br />
N<br />
E<br />
L<br />
IG1<br />
O<br />
Degas<br />
O<br />
IG2<br />
O<br />
On<br />
O<br />
Remote<br />
Figure 4-4<br />
Control Unit Front Panel (with Door Open) Showing Alternate<br />
Cathode Switch.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
4-7