4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for Operation<br />

4.2.1 Pressure Indication During Degas<br />

An approximate indication of pressure within the <strong>Stabil</strong>-<strong>Ion</strong> Gauge is provided on the digital display<br />

during degas. Observing this pressure indication rise, peak out and then fall is an excellent means<br />

of determining the optimum time of degas. The indication may eventually be unrealistically low due<br />

to a reverse x-ray current effect or temporary gauge pumping. Degassing purposely changes surface<br />

conditions in the gauge. After degassing, permit sufficient time for steady state conditions to be<br />

established before attempting to make measurements.<br />

4.2.2 Degas Power Timer<br />

The degas power timer is factory preset for 10 minutes. See Section 2.1.7 on page 2-14 to change<br />

this time.<br />

4.2.3 Degas Power Limiter<br />

For convenience and improved safety, degas power is automatically reduced as gas is liberated to<br />

prevent excessive pressure rise and help reduce the possibility of a gas discharge forming. As the<br />

electrodes clean up, power is increased to a maximum of 40 W. This feature limits the maximum<br />

pressure indication to about 5 x 10 -5 Torr during degas.<br />

4.2.4 Gauge Bakeout<br />

Applying an external heat source to the vacuum system and to the gauge envelope can speed the<br />

removal of contamination from the interior surfaces. Bakeout temperatures between 150 °C and<br />

300 °C are typically adequate. Do not exceed 450 °C at the <strong>Stabil</strong>-<strong>Ion</strong> Gauge envelope. Do not<br />

exceed 150 °C at the cable and connector.<br />

4.3 Alternate Cathode Operation<br />

The cathode coating, which is utilized in the <strong>Stabil</strong>-<strong>Ion</strong> Gauge to provide electron emission at<br />

relatively low cathode temperatures, has a finite operating life—typically two to three years at low<br />

pressure. <strong>Ion</strong> bombardment at higher pressures tends to erode the coating faster than at lower<br />

pressures. <strong>Stabil</strong>-<strong>Ion</strong> Gauges have a second independent cathode for use when the first cathode<br />

finally tires and does not emit sufficiently. Note, the Controller will automatically turn the IG off if<br />

the cathode has insufficient emission either due to contamination of the cathode or long use.<br />

RS-232<br />

ELECTROMETER<br />

CONVECTRON<br />

IG-1 AUTO<br />

PROCESS CONTROL<br />

Off SP On<br />

Data Port<br />

IG2 Scale<br />

IG1<br />

Factor<br />

Adjust<br />

Pressure<br />

Range<br />

L HL H<br />

Cathode<br />

1 21 2<br />

IG1 IG2<br />

OFF SET<br />

ATM<br />

VAC<br />

IG-2 AUTO<br />

OFF SET<br />

ATM<br />

VAC<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

3<br />

▲<br />

Set<br />

▲<br />

C<br />

H<br />

A<br />

N<br />

N<br />

E<br />

L<br />

IG1<br />

O<br />

Degas<br />

O<br />

IG2<br />

O<br />

On<br />

O<br />

Remote<br />

Figure 4-4<br />

Control Unit Front Panel (with Door Open) Showing Alternate<br />

Cathode Switch.<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />

4-7

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!