21.01.2014 Views

Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...

Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...

Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

6 Service and Maintenance<br />

6.5 Electrometer Minimum Calibration Adjustment<br />

This test and adjustment should only be done with the cover on and securely fastened to the<br />

Control Unit. This adjustment is used to cancel residual input currents so that the logarithmic<br />

electrometer amplifier is accurate in its lowest decades. It performs the same function as the zero<br />

adjustment on a conventional electrometer even though there is no zero reading on a log amplifier.<br />

It is factory set by calibrating the pressure readout to correspond to a very small collector input<br />

current. It can be checked in the field without a calibration source using the following procedure:<br />

1. Verify that the scale factor is adjusted to 1.00.<br />

2. Operate a <strong>Stabil</strong>-<strong>Ion</strong> Gauge with 4 mA emission (low pressure range).<br />

3. Disconnect the collector lead from the rear panel of the electrometer module.<br />

4. Observe the display for 10 seconds as it settles to a<br />

minimum reading. This reading should be between<br />

2.50 x 10 -11 and 2.75 x 10 -11 Torr.<br />

5. Locate the MIN. ADJ control on the rear panel of the<br />

Electrometer Module. See Figure 6-3.<br />

6. Generate a display between 2.50 x 10 -11 and 2.75 x 10 -11<br />

Torr by making very small adjustments of less than 1/4 turn<br />

at a time. It will be difficult to achieve a stable reading<br />

because of electronic noise at very low input currents. The<br />

optimum point is where the control causes the display to<br />

just switch to its minimum reading of 2.50 x 10 -11 Torr.<br />

IG1<br />

IG2<br />

ELECTROMETER<br />

COL.<br />

1 19<br />

R<br />

E<br />

M<br />

O<br />

T<br />

E<br />

9 26<br />

COL.<br />

MIN.ADJ.<br />

ANA.<br />

OUT<br />

Figure 6-3<br />

Rear View, Electrometer Module<br />

Showing MIN. ADJ Control.<br />

6.6 Flashing Degas Light<br />

If the DEGAS LED flashes when degassing with cathode 1 selected, but stays on continuously with<br />

cathode 2 selected, this is an indication that only cathode 1 is in poor condition and will eventually<br />

fail to maintain normal operating emission. When this occurs, the IG will fail to stay on with<br />

cathode 1 selected, and you will need to switch to cathode 2 for normal operation. If the same<br />

condition occurs after you have switched to cathode 2, it may indicate that the gauge needs to be<br />

replaced, or that there is a contamination source affecting both cathodes. See Step 6. on page 5-2<br />

for more discussion.<br />

6-6<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!