Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
6 Service and Maintenance<br />
6.5 Electrometer Minimum Calibration Adjustment<br />
This test and adjustment should only be done with the cover on and securely fastened to the<br />
Control Unit. This adjustment is used to cancel residual input currents so that the logarithmic<br />
electrometer amplifier is accurate in its lowest decades. It performs the same function as the zero<br />
adjustment on a conventional electrometer even though there is no zero reading on a log amplifier.<br />
It is factory set by calibrating the pressure readout to correspond to a very small collector input<br />
current. It can be checked in the field without a calibration source using the following procedure:<br />
1. Verify that the scale factor is adjusted to 1.00.<br />
2. Operate a <strong>Stabil</strong>-<strong>Ion</strong> Gauge with 4 mA emission (low pressure range).<br />
3. Disconnect the collector lead from the rear panel of the electrometer module.<br />
4. Observe the display for 10 seconds as it settles to a<br />
minimum reading. This reading should be between<br />
2.50 x 10 -11 and 2.75 x 10 -11 Torr.<br />
5. Locate the MIN. ADJ control on the rear panel of the<br />
Electrometer Module. See Figure 6-3.<br />
6. Generate a display between 2.50 x 10 -11 and 2.75 x 10 -11<br />
Torr by making very small adjustments of less than 1/4 turn<br />
at a time. It will be difficult to achieve a stable reading<br />
because of electronic noise at very low input currents. The<br />
optimum point is where the control causes the display to<br />
just switch to its minimum reading of 2.50 x 10 -11 Torr.<br />
IG1<br />
IG2<br />
ELECTROMETER<br />
COL.<br />
1 19<br />
R<br />
E<br />
M<br />
O<br />
T<br />
E<br />
9 26<br />
COL.<br />
MIN.ADJ.<br />
ANA.<br />
OUT<br />
Figure 6-3<br />
Rear View, Electrometer Module<br />
Showing MIN. ADJ Control.<br />
6.6 Flashing Degas Light<br />
If the DEGAS LED flashes when degassing with cathode 1 selected, but stays on continuously with<br />
cathode 2 selected, this is an indication that only cathode 1 is in poor condition and will eventually<br />
fail to maintain normal operating emission. When this occurs, the IG will fail to stay on with<br />
cathode 1 selected, and you will need to switch to cathode 2 for normal operation. If the same<br />
condition occurs after you have switched to cathode 2, it may indicate that the gauge needs to be<br />
replaced, or that there is a contamination source affecting both cathodes. See Step 6. on page 5-2<br />
for more discussion.<br />
6-6<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001