6 Service and Maintenance<br />

9. Carefully reconnect the bus ribbon connectors.<br />

10. Select appropriate switch settings. See Chapter 2.<br />

11. Replace the top cover as directed in Section 2.1.8 on page 2-14.<br />

6.14 Troubleshooting Flow Charts<br />

Table 6-8<br />

Troubleshooting Flow Charts.<br />

Fig. No. Description Fig. No. Description<br />

Figure 6-7 Power Turn On. Figure 6-10 Convectron Gauge Display.<br />

Figure 6-8 <strong>Stabil</strong>-<strong>Ion</strong> Gauge Turn On. Figure 6-11 RS-232/IEEE 488 Interface.<br />

Figure 6-9<br />

<strong>Stabil</strong>-<strong>Ion</strong> Gauge Display.<br />

<strong>360</strong> TURN ON<br />

SEQUENCE BEGIN<br />

Set switches as follows:<br />

Scale factor: center off<br />

IG1 and IG2 auto: off<br />

POWER SWITCH<br />

ON<br />

REPLACE FUSE<br />

REPLACE CORD<br />

TOGGLE<br />

CIRCUIT BREAKER<br />

YES YES YES<br />

IS<br />

POWER SWITCH<br />

LAMP ON?<br />

NO<br />

CHECK AC<br />

LINE FUSE<br />

IS<br />

FUSE<br />

BLOWN?<br />

NO<br />

CHECK AC<br />

LINE CORD<br />

IS<br />

AC LINE CORD<br />

DAMAGED?<br />

NO<br />

CHECK MAIN<br />

CIRCUIT BREAKER<br />

IS<br />

CIRCUIT<br />

BREAKER<br />

TRIPPED?<br />

YES<br />

NO<br />

IS<br />

ION GAUGE<br />

PRESSURE<br />

DISPLAY<br />

BLANK?<br />

NO<br />

YES<br />

ARE<br />

IG GAUGE<br />

#1 AND #2<br />

ON/OFF<br />

LAMPS<br />

OFF?<br />

NO<br />

SET SWITCHES<br />

TO CORRECT<br />

VOLTAGE<br />

REPLACE CABLE<br />

YES<br />

YES<br />

YES<br />

ARE<br />

CONVECTRON<br />

A & B<br />

DISPLAYS<br />

ON?<br />

NO<br />

TURN OFF<br />

POWER SWITCH<br />

CHECK POWER<br />

SUPPLY AC<br />

LINE VOLTAGE<br />

SWITCHES<br />

ARE<br />

SWITCHES<br />

NOT SET<br />

CORRECTLY?<br />

NO<br />

CHECK<br />

CONTROLLER<br />

INTERCONNECT<br />

CABLE<br />

IS<br />

CONTR.<br />

INTERCONNECT<br />

CABLE<br />

DAMAGED?<br />

NO<br />

CALL FACTORY<br />

YES<br />

<strong>360</strong> TURN ON<br />

SEQUENCE FINISH<br />

Figure 6-7<br />

Power Turn On.<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />

6-11