6 Service and Maintenance<br />
9. Carefully reconnect the bus ribbon connectors.<br />
10. Select appropriate switch settings. See Chapter 2.<br />
11. Replace the top cover as directed in Section 2.1.8 on page 2-14.<br />
6.14 Troubleshooting Flow Charts<br />
Table 6-8<br />
Troubleshooting Flow Charts.<br />
Fig. No. Description Fig. No. Description<br />
Figure 6-7 Power Turn On. Figure 6-10 Convectron Gauge Display.<br />
Figure 6-8 <strong>Stabil</strong>-<strong>Ion</strong> Gauge Turn On. Figure 6-11 RS-232/IEEE 488 Interface.<br />
Figure 6-9<br />
<strong>Stabil</strong>-<strong>Ion</strong> Gauge Display.<br />
<strong>360</strong> TURN ON<br />
SEQUENCE BEGIN<br />
Set switches as follows:<br />
Scale factor: center off<br />
IG1 and IG2 auto: off<br />
POWER SWITCH<br />
ON<br />
REPLACE FUSE<br />
REPLACE CORD<br />
TOGGLE<br />
CIRCUIT BREAKER<br />
YES YES YES<br />
IS<br />
POWER SWITCH<br />
LAMP ON?<br />
NO<br />
CHECK AC<br />
LINE FUSE<br />
IS<br />
FUSE<br />
BLOWN?<br />
NO<br />
CHECK AC<br />
LINE CORD<br />
IS<br />
AC LINE CORD<br />
DAMAGED?<br />
NO<br />
CHECK MAIN<br />
CIRCUIT BREAKER<br />
IS<br />
CIRCUIT<br />
BREAKER<br />
TRIPPED?<br />
YES<br />
NO<br />
IS<br />
ION GAUGE<br />
PRESSURE<br />
DISPLAY<br />
BLANK?<br />
NO<br />
YES<br />
ARE<br />
IG GAUGE<br />
#1 AND #2<br />
ON/OFF<br />
LAMPS<br />
OFF?<br />
NO<br />
SET SWITCHES<br />
TO CORRECT<br />
VOLTAGE<br />
REPLACE CABLE<br />
YES<br />
YES<br />
YES<br />
ARE<br />
CONVECTRON<br />
A & B<br />
DISPLAYS<br />
ON?<br />
NO<br />
TURN OFF<br />
POWER SWITCH<br />
CHECK POWER<br />
SUPPLY AC<br />
LINE VOLTAGE<br />
SWITCHES<br />
ARE<br />
SWITCHES<br />
NOT SET<br />
CORRECTLY?<br />
NO<br />
CHECK<br />
CONTROLLER<br />
INTERCONNECT<br />
CABLE<br />
IS<br />
CONTR.<br />
INTERCONNECT<br />
CABLE<br />
DAMAGED?<br />
NO<br />
CALL FACTORY<br />
YES<br />
<strong>360</strong> TURN ON<br />
SEQUENCE FINISH<br />
Figure 6-7<br />
Power Turn On.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />