6 Service and Maintenance<br />

Symptom<br />

Green +15 LED out.<br />

Green -15 LED out.<br />

Green +5 display LED out.<br />

Green +5 logic LED out.<br />

Possible Cause<br />

+15 volt supply faulty (power to analog circuitry and RS-232).<br />

-15 volt supply faulty (power to analog circuitry and RS-232).<br />

+5 volt supply to display LED's faulty.<br />

+5 volt logic supply faulty.<br />

6.3 Fault Relay<br />

When supplied with the available factory installed <strong>Stabil</strong>-<strong>Ion</strong> Remote<br />

Control (Cat. No. <strong>360</strong>104), the <strong>360</strong> Control Unit is equipped with a “Fault”<br />

relay. The SPDT relay contacts are located on the remote output connector<br />

(See Figure 6-1). The relay is energized when no fault condition is detected.<br />

The relay is de-energized when:<br />

■<br />

■<br />

■<br />

■<br />

■<br />

There is a fault condition detected by one of the microprocessors.<br />

There is a microprocessor failure causing the watchdog timer to reset.<br />

Power is lost due to Power Supply failure.<br />

Power is removed from the instrument.<br />

The power switch is turned off.<br />

Figure 6-1<br />

Control Board Top<br />

View Showing Green<br />

Power Indicator LED’s.<br />

The source of the fault signal can be determined by observing the fault indicating red LED's and the<br />

Power Supply monitor's green LED's located beneath the Control Unit's cover (see Figure 1-8 on<br />

page 1-6). A list of specific causes are described in Table 6-2.<br />

Table 6-2<br />

Fault Cause/Action.<br />

Cause<br />

Microprocessor reset occurring on some module (probable<br />

circuit failure).<br />

Unplugged Convectron Gauge.<br />

Sensor broken in Convectron Gauge.<br />

Action<br />

Cycle power.<br />

Call Qualified Service Technician.<br />

Plug in gauge.<br />

Replace gauge.<br />

Convectron Gauge zero badly out of calibration. Adjust zero calibration. See Section 4.9 on page 4-24.<br />

Checksum failure on Process Control board.<br />

Cycle power and check all setpoints.<br />

6-4<br />