6 Service and Maintenance<br />
Symptom<br />
Green +15 LED out.<br />
Green -15 LED out.<br />
Green +5 display LED out.<br />
Green +5 logic LED out.<br />
Possible Cause<br />
+15 volt supply faulty (power to analog circuitry and RS-232).<br />
-15 volt supply faulty (power to analog circuitry and RS-232).<br />
+5 volt supply to display LED's faulty.<br />
+5 volt logic supply faulty.<br />
6.3 Fault Relay<br />
When supplied with the available factory installed <strong>Stabil</strong>-<strong>Ion</strong> Remote<br />
Control (Cat. No. <strong>360</strong>104), the <strong>360</strong> Control Unit is equipped with a “Fault”<br />
relay. The SPDT relay contacts are located on the remote output connector<br />
(See Figure 6-1). The relay is energized when no fault condition is detected.<br />
The relay is de-energized when:<br />
■<br />
■<br />
■<br />
■<br />
■<br />
There is a fault condition detected by one of the microprocessors.<br />
There is a microprocessor failure causing the watchdog timer to reset.<br />
Power is lost due to Power Supply failure.<br />
Power is removed from the instrument.<br />
The power switch is turned off.<br />
Figure 6-1<br />
Control Board Top<br />
View Showing Green<br />
Power Indicator LED’s.<br />
The source of the fault signal can be determined by observing the fault indicating red LED's and the<br />
Power Supply monitor's green LED's located beneath the Control Unit's cover (see Figure 1-8 on<br />
page 1-6). A list of specific causes are described in Table 6-2.<br />
Table 6-2<br />
Fault Cause/Action.<br />
Cause<br />
Microprocessor reset occurring on some module (probable<br />
circuit failure).<br />
Unplugged Convectron Gauge.<br />
Sensor broken in Convectron Gauge.<br />
Action<br />
Cycle power.<br />
Call Qualified Service Technician.<br />
Plug in gauge.<br />
Replace gauge.<br />
Convectron Gauge zero badly out of calibration. Adjust zero calibration. See Section 4.9 on page 4-24.<br />
Checksum failure on Process Control board.<br />
Cycle power and check all setpoints.<br />