Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for Operation<br />
Table 4-1 Optimum Pressure Range Settings for N 2 .<br />
Switch Setting<br />
Pressure Extreme<br />
(Torr) Emission Overpressure Setpoint<br />
H - High Range above 10 -4 0.1 mA 2x10 -2 Torr, or mbar; 2 Pa<br />
L - Low Range below 10 -7 4 mA 5x10 -4 Torr, or mbar; 5 x 10 -2 Pa<br />
For example, if pressures in your application will range from 1 x 10 -2 to say 5 x 10 -8 Torr, leave<br />
the range switch on H. If much of the time the system pressure will be below approximately<br />
1 x 10 -4 Torr, leave the switch on L. We suggest you do not change the range switch frequently<br />
because there is a significant stabilization time to obtain steady state conditions.<br />
NOTE: When the cathode is first turned on, or has been exposed to atmospheric pressure, it is<br />
normal for a relatively large burst of gas to be released momentarily. Using the H range will help<br />
prevent a nuisance shutdown from this burst.<br />
5. When the pressure in the vicinity of IG1 is less than the overpressure setpoint listed above, as<br />
indicated by a Convectron Gauge or another gauge, turn on IG1. See Section 4.8 on page 4-23<br />
for automatic operation.<br />
6. The N 2 equivalent pressure within IG1 will be displayed in the pressure units you have<br />
specified (Torr, mbar or Pa). See Section 2.1.3 on page 2-4 to change pressure units.<br />
7. Steps 5 and 6 also apply for IG2 if your <strong>Stabil</strong>-<strong>Ion</strong> System is equipped with the Dual <strong>Stabil</strong>-<strong>Ion</strong><br />
Operation module.<br />
4.1.1 Alternate ON/OFF Gauge Control<br />
The <strong>Stabil</strong>-<strong>Ion</strong> Gauges may be turned on and off in 4 ways:<br />
■<br />
Using the front panel push buttons, IG1 and IG2.<br />
■ Remotely with the available Remote Control module, see Section 3.7 on page 3-9.<br />
■<br />
■<br />
Automatically using the auto on function of the Convectron Gauge module. See Section<br />
4.8 on page 4-23.<br />
Using the RS-232 or IEEE 488 Computer Interface modules. See Section 4.12 on page<br />
4-27 or Section 4.13 on page 4-31.<br />
In addition, the <strong>Stabil</strong>-<strong>Ion</strong> Gauge will be automatically turned off by excessive pressure.<br />
4.1.2 Gas Scale Factor Adjustment<br />
If highly accurate measurement of pressure of gases is required, we recommend calibration for the<br />
specific gas and system to be used. See Section 4.4 on page 4-8. The standard calibration is for N 2 .<br />
Different gases have different ionization probabilities. If a <strong>Stabil</strong>-<strong>Ion</strong> Gauge and Controller are<br />
calibrated for N 2 , and if argon is the predominant gas in the system, setting the scale factor at 7.75 x<br />
10 -1 as shown in Table 4-2 on page 4-5 causes the pressure indications for N 2 to be multiplied by<br />
0.775, thus providing correct pressure indications for argon over most of the pressure range. At<br />
pressures above about 1 x 10 -5 Torr this method may not provide accurate pressure indications.<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001<br />
4-3